Self-biased charge sampling amplifier in 90nm CMOS for medical ultrasound imaging

Published: 11 March 2007 Publication History

Abstract

In this paper, we present the analysis and design of a self-biased single-ended charge sampling amplifier (CSA) in 90nm CMOS for catheter based intravenous ultrasound imaging applications. The proposed CSA is based on a 1V single-ended CMOS inverter-based cascode amplifier. The amplifier achieves a DC gain of 43.7 dB and a unity gain frequency of 1.37 GHz at a power consumption of 385 μW at 37°C - nominal temperature of human body with typical-typical (TT) models defined in the 90nm CMOS technology used for this design. Performance of the self-biased CSA is studied by connecting a single Capacitive Micro machined Ultrasound Transducer (CMUT) to it.

References

[1]
Wygant, I. O., et al. Integrated Ultrasonic Imaging Systems Based on CMUT Arrays: Recent Progress. IEEE Ultasonics Symposium, 2004, pp. 391--394.
[2]
Wygant, I. O., et al. An Endoscopic Imaging System Based on a Two-Dimensional CMUT array: Real Time Imaging Results. IEEE Ultasonics Symposium, 2005, pp. 792--795.
[3]
Annema, A. J., et al. Analog Circuits in Ultra-Deep-Submicron CMOS. IEEE Journal of Solid-State Circuits, 2005, Vol. 40, no. 1, pp. 132--143.
[4]
Taur, Y. CMOS design near the limit of scaling. IBM J. Res. & Dev., 2002, vol. 46, no. 2/3, pp. 213--222.
[5]
Buss, D. D. Technology in the internet age. IEEE Int. Solid-State Circuits Conf. (ISSCC) Dig. Tech. Papers, 2002, pp. 18--21.
[6]
Annema, A. J., et al. H. Designing outside rails constraints. IEEE Int. Solid-State Circuits Conf. (ISSCC) Dig. Tech. Papers, 2004, pp. 134--135.
[7]
http://www.iet.ntnu.no/projects/smida/
[8]
Gang Xu and Jiren Yuan. Performance analysis of general charge sampling. IEEE Trans. on Circuits and Systems-II: Express Briefs, February 2005, Vol. 52, No. 2, pp. 107--111.
[9]
Gang Xu and Jiren Yuan. Comparison of charge sampling and voltage sampling. MWSCAS-2000.
[10]
Gang Xu and Jiren Yuan. A Low voltage high-speed sampling technique. Proceedings 4th International conference on ASIC, 2001, pp. 228--231.
[11]
Linga Reddy Cenkeramaddi and Trond Ytterdal. Jitter analysis of general Charge Sampling Amplifiers. IEEE ISCAS, 2006, pp. 5267--5270.
[12]
David A Johns and Ken Martin. Analog Integrated Circuit Design. 1997 John Wiley & Sons.
[13]
Kjersti Midtbø and Arne Rønnekleiv. Fabrication and Characterization of CMUTs realized by wafer bonding. IEEE Ultrasonics Symposium 2006.

Cited By

View all
  • (2018)Feedback Biasing Based Adjustable Gain Ultrasound Preamplifier for CMUTs in 45nm CMOS2018 31st International Conference on VLSI Design and 2018 17th International Conference on Embedded Systems (VLSID)10.1109/VLSID.2018.63(204-207)Online publication date: Jan-2018

Index Terms

  1. Self-biased charge sampling amplifier in 90nm CMOS for medical ultrasound imaging

    Recommendations

    Comments

    Information & Contributors

    Information

    Published In

    cover image ACM Conferences
    GLSVLSI '07: Proceedings of the 17th ACM Great Lakes symposium on VLSI
    March 2007
    626 pages
    ISBN:9781595936059
    DOI:10.1145/1228784
    Permission to make digital or hard copies of all or part of this work for personal or classroom use is granted without fee provided that copies are not made or distributed for profit or commercial advantage and that copies bear this notice and the full citation on the first page. Copyrights for components of this work owned by others than ACM must be honored. Abstracting with credit is permitted. To copy otherwise, or republish, to post on servers or to redistribute to lists, requires prior specific permission and/or a fee. Request permissions from [email protected]

    Sponsors

    Publisher

    Association for Computing Machinery

    New York, NY, United States

    Publication History

    Published: 11 March 2007

    Permissions

    Request permissions for this article.

    Check for updates

    Author Tags

    1. CMUT-CMOS
    2. CSA
    3. analog front-end for CMUTs
    4. charge sampling
    5. sampling

    Qualifiers

    • Article

    Conference

    GLSVLSI07
    Sponsor:
    GLSVLSI07: Great Lakes Symposium on VLSI 2007
    March 11 - 13, 2007
    Stresa-Lago Maggiore, Italy

    Acceptance Rates

    Overall Acceptance Rate 312 of 1,156 submissions, 27%

    Upcoming Conference

    GLSVLSI '25
    Great Lakes Symposium on VLSI 2025
    June 30 - July 2, 2025
    New Orleans , LA , USA

    Contributors

    Other Metrics

    Bibliometrics & Citations

    Bibliometrics

    Article Metrics

    • Downloads (Last 12 months)0
    • Downloads (Last 6 weeks)0
    Reflects downloads up to 27 Jan 2025

    Other Metrics

    Citations

    Cited By

    View all
    • (2018)Feedback Biasing Based Adjustable Gain Ultrasound Preamplifier for CMUTs in 45nm CMOS2018 31st International Conference on VLSI Design and 2018 17th International Conference on Embedded Systems (VLSID)10.1109/VLSID.2018.63(204-207)Online publication date: Jan-2018

    View Options

    Login options

    View options

    PDF

    View or Download as a PDF file.

    PDF

    eReader

    View online with eReader.

    eReader

    Figures

    Tables

    Media

    Share

    Share

    Share this Publication link

    Share on social media